Liquiline System CA80TN
Operation
Hauser
91
•
Start date
= 01.01.2020
•
Start time
= 12:00:00
•
Signal duration
= 05:00:00
•
Period
= 24:00:00
3. Create
Formula
mathematical function.
Menu
/
Setup
/
Additional functions
/
Mathematical functions
• MF1: Formula
•
Tracking
=
On
•
Source A
=
Time switch 1
•
Source B
=
Time switch 2
•
Formula
= 18.0 + 3*NUM(A) + 2*NUM(B)
Explanation: NUM converts the logical value to a numerical value and thus enables
multiplication.
• 3*NUM(A) results in the value 3.0 from 08:00 to 17:00 and 0.0 outside of this time
period.
• 2*NUM(B) results in the value 2.0 from 12:00 to 17:00 and 0.0 outside of this time
period.
The formula thus yields one of these analog values depending on the time: 18.0, 21.0 or
23.0. This analog value can be used as a set point for a temperature controller.
Example 2: Time-based condition
A pump should switch on (via a relay) for 10 minutes every 2 hours. This only applies if
the pH value is lower than 4.0.
1. Program
Time switch 1
:
•
Start date
= 01.01.2020
•
Start time
= 00:00:00
•
Signal duration
= 00:10:00
•
Period
= 02:00:00
2. Create
Formula
mathematical function.
Menu
/
Setup
/
Additional functions
/
Mathematical functions
• MF1: Formula
•
Tracking
=
On
•
Source A
=
Time switch 1
•
Source B
= pH value of a Memosens pH input
•
Formula
= A AND(B<4.0)
3. Use the formula as the data source of a relay.
Menu
/
Setup
/
Outputs
/
Relay
[x:y]
•
Function
=
Formula
•
Operating mode
=
Static signal
•
Source of data
=
MF1: Formula
The formula yields a logical value (TRUE or FALSE) and is thus suitable to trigger a relay
directly in the static operating mode.
Time switch 1
provides a TRUE value for 10 minutes
every 2 hours but only if the pH value has dropped to below 4 at the same time.