Mounting
Liquiline System CA80TN
16
Hauser
‣
Protect the device against mechanical vibrations.
‣
Protect the device against corrosive gases, e.g. hydrogen sulfide (H
2
S) .
‣
Make sure to pay attention to the maximum height difference and the maximum
distance from the sampling point.
‣
Ensure that the unit can drain freely, without any siphoning effects.
‣
Make sure air can circulate freely at the front of the housing.
‣
Open analyzers (i.e. analyzers that are supplied without a door) may only be erected in
closed areas or in a protective cabinet or similar facility.
5.1.3
Spacing requirements when mounting
Spacing required for installing analyzer
≥330(12.99)
≥
100 (3.94)
A0036774
12
Minimum spacing required for mounting.
Engineering unit mm (in).
≤116 °
A0036775
13
Maximum opening angle
Spacing required for installing wall-mount version
750 (29.53)
120 (4.72)
Ø 11 (0.43)
Ø 10.7 (0.42)
A0036779
14
Holder unit dimensions. Engineering unit mm (in)