Maintenance
Liquiline System CA80TN
138
Hauser
4. Select
Menu
/
Operation
/
Maintenance
/
Decommissioning
/
Empty hoses
to remove
the liquid from the hoses.
The hoses are automatically purged with air and emptied. This process takes
several minutes.
Following the successful completion of the action, the following message appears
The action was successful
.
12.3.3 Rinsing the analyzer
1. Open all the reagent bottles and carefully remove the hoses together with the covers.
Then close the bottles firmly.
2. Dry the ends of the hoses with a clean paper towel.
3. Immerse all the hoses into a beaker with approx. 200 ml (6.76 fl oz) of distilled
water in order to rinse the system with water.
4. Select
Menu
/
Operation
/
Maintenance
/
Decommissioning
/
Rinse with water
.
The system is rinsed automatically and emptied. This process takes approx. 8
minutes.
Following the successful completion of the action, the following message appears
The action was successful
.
5. Once the action is finished, remove the hoses from the beaker and dry them with a
clean paper towel.
12.3.4 Emptying the hoses again
1.
Menu
/
Operation
/
Maintenance
/
Decommissioning
/
Empty hoses
must be selected
to remove the liquid from the hoses.
The hoses are automatically purged with air and emptied. This process takes
several minutes.
2. Wait until this action has finished. Following the successful completion of the action,
the following message appears
The action was successful
.
3. The analyzer can now be disconnected from the mains supply.