Technical data
Liquiline System CA80TN
156
Hauser
Electromagnetic
compatibility
Interference emission and interference immunity as per EN 61326-1:2013, Class A for
Industry
Electrical safety
According to EN/IEC 61010-1:2010, Class I equipment
Low voltage: overvoltage category II
For installations up to 2000 m (6500 ft) above MSL
Pollution degree
Pollution level 2
15.9 Process
Sample temperature
4 to 40 °C (39 to 104 °F)
Consistency of the sample
Low solids content, aqueous, homogenized
Sample supply
Analyzer (without optional Y strainer):
Unpressurized
Y strainer (optional):
• Permitted pressure range: max. 4 bar (58.01 psi)
• Flow: ensure sufficient flow for a Y strainer that is completely filled
15.10 Mechanical construction
Dimensions
Materials
Parts not in contact with medium
Cabinet version, exterior cover
Plastic ASA+PC
Open installation, exterior cover
Cabinet version, interior lining
Plastic PP
Open installation, interior lining
Window
Shatterproof glass, coated
Reagent container
Plastic PP
Base, analyzer stand
Powder-coated sheet steel
Parts in contact with medium
• Manifold
• Valve seals
• Pinch valve
• Plastic ETFE
• Plastic FKM
• Plastic PP and PCTFE
6)
Sufficient mains quality is required to operate the product as intended.
7)
The greater the flow the greater the self-cleaning effect of the suction line. Preferred: > 1 m³/h