Liquiline System CA80TN
Maintenance
Hauser
135
6. Perform photometer test:
DIAG
/
System test
/
Analyzer
/
Photometer
/
Photometer
test
/
Start for 5 min
.
7.
Current intensity 1
: Is the value higher than the value initially noted down?
If yes, cleaning was successful.
8. To conclude, perform a zero point calibration:
Menu
/
Operation
/
Manual operation
/
Determine zero point
.
Cleaning the fan of the cooling module
A0045403
66
Fan of the cooling module (behind the ventilation grille)
1. Check the fan regularly for dirt.
2. In environments with a heavy particle load:
vacuum clean regularly from outside.
12.2.4 Replacing reagents
Removing bottles
1. Stop the sample supply.
2. If you have not already done so,
set the manual mode.
3. Pull the hoses (RB, RK, RN, S1, S0, C) out of the reagent bottles as far as they will go
so that the hose ends are outside of the liquid.
4. Select
Menu
/
Operation
/
Maintenance
/
Bottle change mode
/
Bottle removal
/
Bottle selection
.
5. Select the bottles that you want to remove and press
OK
to confirm.
6. Select
Confirm bottles removed
.
Replacing bottles
1. Replace the removed bottles with fresh reagent .
Reagents must be prepared according to the instructions for mixing a reagent.
2. Select
Menu
/
Operation
/
Maintenance
/
Bottle change mode
/
Bottle insertion
/
Bottle selection
.
3. Select all the bottles that you have replaced and press
OK
to confirm.
4. Select
Confirm bottles inserted
.
5. If bottle level monitoring is activated (
Menu
/
Setup
/
Analyzer
/
Extended setup
/
Diagnostics settings
/
Bottles
), you must reset this under
Menu
/
Operation
/
Maintenance
/
Bottle change mode
/
Bottle insertion
/
Reset filling levels
.