Liquiline System CA80TN
Maintenance
Hauser
137
4. Reset the operating hours counter for the mats:
Menu
/
Operation
/
Operating time
filter mats
/
Reset
.
12.3 Decommissioning
L
CAUTION
Activities while the analyzer is in operation
Risk of injury and infection from medium!
‣
Before you release any hoses, make sure that no actions, such as the pumping of
sample, are currently running or are due to start shortly.
‣
Wear protective clothing, goggles and gloves or take other suitable measures to protect
yourself.
‣
Wipe up any spilt reagent with a disposable tissue and rinse with clear water. Then dry
the cleaned areas with a cloth.
NOTICE
Do not operate the analyzer for longer than 5 days without taking it out of service
properly!
Can result in damage to the device.
‣
Take the analyzer out of service as described.
To take out of service, proceed as follows:
1. Set the manual mode
2. Empty the hoses
3. Rinse the analyzer
4. Empty the hoses again
12.3.1 Setting the manual mode
Initial situation: The device is still in the automatic mode and you now want to take the
device out of service.
1. Press the
MODE
soft key.
The current mode is displayed:
Automatic
2. Select an action:
Manual mode
The device changes to the manual mode.
3. Wait until all operations have stopped.
4.
Alternatively:
Stop actions currently running under:
Menu
/
Operation
/
Manual operation
.
12.3.2 Emptying the hoses
1. Pull the hoses (RB, RK, C, S1, S0) out of the reagent bottles as far as they will go so
that the hose ends are outside of the liquid.
2. Only for version CA80TN-HR:
Also pull the REF hose out of the liquid as far as it will go.
3. Pull the sample hose (P) out of the bottle, and if the optional dilution module is
installed pull the hose to the water for the dilution module (W) out of the bottle, so
that the hose end is outside of the liquid.