Maintenance
Liquiline System CA80TN
132
Hauser
12.2 Maintenance tasks
12.2.1 Software menu Maintenance
Menu/Operation/Maintenance
• Bottle replacement
If level monitoring is activated: specify when bottles are replaced.
• Filter mats replacement
Operating time counter, can be reset here. Replace filter mats: → 136
• Pump hose replacement
Operating time counter, can be reset here. Replace hose: → 136
• Valve hose replacement
Operating time counter, can be reset here. Replace hose: → 136
• Commissioning
Automatic commissioning, no recurring maintenance tasks → 43
• Decommissioning
• Reactor maintenance
• Emptying reactor
• Flushing reactor
12.2.2 Setting the manual mode
Initial situation: The device is still in the automatic mode and you want to perform
maintenance work.
1. Press the
MODE
soft key.
The current mode is displayed:
Automatic
2. Select an action:
Manual mode
The device changes to the manual mode.
3. Wait until all operations have stopped.
4.
Alternatively:
Stop actions currently running under:
Menu
/
Operation
/
Manual operation
.
12.2.3 Cleaning
L
CAUTION
Risk of injury from reagents escaping
‣
Before replacing, always rinse the system first.
Cleaning the housing
‣
Clean the front of the housing using commercially available cleaning agents only.
The front of the housing is resistant to the following in accordance with DIN 42 115:
• Ethanol (for a short time)
• Diluted acids (max. 2% HCl)
• Diluted bases (max. 3% NaOH)
• Soap-based household cleaning agents
NOTICE
Cleaning agents not permitted
Damage to the housing surface or housing seal
‣
Never use concentrated mineral acids or alkaline solutions for cleaning.
‣
Never use organic cleaners such as acetone, benzyl alcohol, methanol, methylene
chloride, xylene or concentrated glycerol cleaner.
‣
Never use high-pressure steam for cleaning.