Liquiline System CA80TN
Diagnostics and troubleshooting
Hauser
117
No.
Message
Factory settings
Tests or remedial action
S
D
F
541
Parameter load ok
M
On
Off
Configuration successfully loaded
542
Parameter load fail
M
On
Off
Loading of configuration has failed
‣
Repeat.
543
Parameter load abort
M
On
Off
Configuration loading aborted
544
Parameter reset ok
M
On
Off
Factory default successful
545
Parameter reset fail
M
On
Off
Setting of device configuration to factory setting
has failed
565
Configuration
M
On
Off
Invalid configuration of sample preparation
system
1.
In
Setup
/
Sample preparation
, check the
number of channels used, their operating
mode and type of installation.
2.
Check the permitted combinations of
sample preparation and analyzers, see the
Operating Instructions for the sample
preparation system.
714
Filter mats change
M
On
Off
Filter mat change necessary
The limit value for operating hours has been
exceeded
‣
Replace the filter mats and reset the
operating hours counter in the Diagnostics
menu.
715
Calibration expired
M
On
Off
The validity of the last calibration has expired.
The date of the last calibration is too far in the
past. Measurement can still take place.
Possible reasons:
Manual intervention has prevented automatic
calibration
1.
Calibrate analyzer manually.
2.
Check device configuration.
716
Calibration expired
S
On
Off
Calibration has failed or is not reliable
Possible reasons:
Stability criteria are not met
1.
Check the configuration and perform a
manual calibration again.
2.
Contact Hauser Service.
717
Photometer defective
F
On
On
Photometer defective
Possible reasons:
• No lamp voltage
• No lamp power
‣
Contact Hauser Service.