Commissioning
Liquiline System CA80TN
42
Hauser
Locking access to the configuration
‣
Call up:
MENU
/
Exit maintenance level
.
Access is locked again and you cannot change any settings.
9.5
Setting the operating language
Configuring the language
1. Press the soft key:
MENU
.
2. Set your language in the top menu item.
The device can now be operated in your chosen language.
9.6
Configuring the measuring device
9.6.1
Basic setup analyzer
Making basic settings
1. Switch to the menu
Setup
/
Basic setup analyzer
.
Make the following settings.
• Device tag
Give your device any name of your choice (max. 32 characters).
• Set date
Correct the set date if necessary.
• Set time
Correct the set time if necessary.
2. Insert the bottles and activate the bottles used in the menu:
Bottle insertion
/
Bottle
selection
.
3. Check the concentration of the calibration standard used:
Calibration
/
Settings
/
Nominal concentration
.
4. Optionally, also change the measuring interval:
Measurement
/
Measuring interval
.
All the other settings can be left in the default factory settings for the time being.
5. Return to the measuring mode: press and hold the soft key for
ESC
for at least one
second.
Your analyzer now works with your general settings. Optionally connected
sensors use the factory settings of the specific sensor type and the individual
calibration settings that were last saved.
If you want to already configure additional input and output parameters in the
Basic setup
analyzer
:
‣
Configure the current outputs, relays, limit switches and device diagnostics with the
following submenus.