Liquiline System CA80TN
Operation
Hauser
59
10.3.1 Extended setup
Measuring value
Menu/Setup/Analyzer/Extended setup/Measuring value
Function
Options
Info
Main value
Selection
N
Factory setting
N
Unit
Selection
• mg/l
• ppm
Factory setting
mg/l
Main value format
Selection
• #
• #.#
• #.##
• #.###
• Auto
Factory setting
#.#
Status for failed measurement
Selection
• Keep last value
• Status bad
Factory setting
Keep last value
Keep last value
Standard practice is to keep the measured value
of the last successful measurement until the next
successful measurement.
To detect an error during measurement, the
NAMUR status signal, the time stamp of the
measurement and the measured value status
must be evaluated.
Status bad
If a measurement cannot be performed due to an
error, the measured value status is set to
Bad
and the measured value to 0. The time stamp is
updated accordingly.
Menu/Setup/Analyzer/Extended setup
Function
Options
Info
Upper range limit
Menu only available for device
versions
CA80TN-HR
Configurable
• At least 10% of the
upper limit value of the
smallest available
measuring range
• The maximum upper
limit value of the largest
available measuring
range (factory setting)
Range of adjustment depends on: Menu/Setup/
Analyzer/Measuring range
‣
Specify the desired value for the measuring
range upper limit.
Directly affects the display:
Menu
/
Setup
/
Analyzer
/
Measuring range
.
Device range
Menu not available for device
versions
CA80TN-HR
Selection
The options available depend on the device
version ordered.
‣
Select the desired measuring range from the
list.