Operation
Liquiline System CA80TN
66
Hauser
Menu/Setup/Analyzer/Calibration
Function
Options
Info
Calibration factor
Read only
Relationship of the measured concentration to
the pre-specified concentration of the calibration
standard.
Settings
Nominal concentration
1.00 to 200.00 ppm
Factory setting
10.00 ppm
For configuring the concentration of the
calibration standard solution.
Depends on the settings under
Menu
/
Setup
/
Analyzer
/
Extended setup
/
Measuring value
/
Main value
Automatic cleaning
Selection
• Off
• On
Factory setting
Off
For specifying whether cleaning will take place
before every calibration/adjustment (only in the
automatic mode).
10.3.5 Cleaning
Menu/Setup/Analyzer/Cleaning
Function
Options
Info
Start condition
Selection
• Immediate
• Date/time
• Disabled
Factory setting
Immediate
Immediate
: Cleaning of the starts immediately.
Date/time
: Cleaning of the starts at the selected
date/time.
Disabled
: Additional cleaning is not enabled.
Date
01.01.1970 to 07.02.2106
Functions only available for:
Start condition
=
Date/time
Time
00:00:00 to 23:59:59
Cleaning interval
0-01 to 90-00 (
DD-hh
)
Factory setting
02-00
DD-hh
For configuring the time interval for (additional)
cleaning.
Next cleaning
Mode
=
Automatic
Read only
Settings
Sample hose cleaning
Selection
• On
• Off
Factory setting
Off