Maintenance
Liquiline System CA80TN
136
Hauser
6. Following the replacement, a zero point calibration must be performed and the
calibration factor must be redetermined:
Menu
/
Operation
/
Manual operation
/
Determine zero point
and /
Determine calibration factor
(in the versions CA80TN-
AA/CA/GR, both actions are provided in one function:
Determine zero point and
calibr. factor
).
7. Go back to the
MODE
/
Continue automatic mode
or
MODE
/
Start automatic mode
to start the normal measurement.
12.2.5 Replacing the hoses
There should not be any more sample left in the dosing unit.
2. Remove the bottles and the bottle tray.
3. Replace the hoses with hoses of the same diameter and length. Attach hose labels to
the new hoses.
4. Tighten the hose glands carefully; leaks can result in bad measured values.
5. Secure the cover and place the bottle tray back into the housing.
6. Connect the bottles to the appropriate covers and hoses.
7. Select
Menu
/
Operation
/
Maintenance
/
Bottle change mode
/
Bottle insertion
/
Bottle selection
.
8. Select all the bottles and press
OK
to confirm.
9. Select
Confirm bottles inserted
.
10. Press the
MODE
soft key and select
Continue automatic mode
or
Start automatic
mode
.
12.2.6 Replacing the filter mats
You require the following parts:
Filter mats (part of the CAV maintenance kit)
A0045364
67
Fan screen on the bottom
1. Open both fan screens and remove them.
The filter mats are located under the screens.
2. Remove the used filter mats and replace them with new ones from the maintenance
kit.
3. Fit the fan screens back on.
Summary of Contents for Liquiline System CA80TN