Biometra TOne
Operation
35
Program parameters
Display
(example)
Description
ΔR
!
(in red)
Adjust the average heating and cooling rate
for the selected step
The maximum heating and cooling rate de-
pends on the device model.
Degrees
55.0 +1.0
Increase or decrease the block temperature
column by column by the temperature incre-
ment [°C] starting from the annealing tem-
perature [°C]
Spreadsheet programming
The currently selected step is highlighted in blue during the programming process.
Name:
Lid:
Preheat Lid:
On
Home
Edit
Insert
Delete
Done
Back
Loops
Step
Block Temp. (°C)
Hold Time
(h:mm:ss)
Go To
Cycles
∆T(°C)
∆t(s)
∆R(°C/s)
Three-step
dd.mm.yyyy hh:mm
Program Runtime:
1h 15m 10s
User:
John Doe
Fig. 14
Spreadsheet programming
Program parameters
Display
(example)
Description
Name:
Three-step
Define the program name
Lid:
99 °C
Define the temperature of the heated lid
(30 to 110 °C)
Preheat Lid:
On
Enable or disable the preheating of the
heated lid before the actual PCR run
Step
1
Define at least the block temperature and the
holding time for each step. All other parame-
ters are optional.
For more than 6 steps, use the arrow keys to
navigate through the displayed program
Block Temp. (°C)
95.0
Define a block temperature (3 to 99 °C
±0.1 °C) for each step
Hold Time (h:mm:ss)
0:05:00
Define a holding time during which the block
temperature is maintained
(≤9 h 59 min 59 s)
Use
∞
to program a pause