Biometra TOne
Operation
45
Display
Examples
Description
Name
Three-step
Program name
¡
For programs with gradient: Display
(Degrees)
behind the program names
Estimated End
17:49:46
Expected completion time of the program
Remain. Time
01h 26m 11s
Expected remaining run time
Block Temp.
59.5°C
Current block temperature
¡
For programs with gradient: Display of the
lowest and highest block temperature,
separated by the icon
Lid
99.7°C
Current temperature of the heated lid
Step
2/6
Preheating
Current step in the step sequence
Display during the preheating process of the lid
and the subsequent equilibration (40 s)
5.8.1
Understanding the graphical, spreadsheet and gradient view
Understanding the graphical
view
During a program run, the current step is highlighted in yellow. The graphical view di-
vides each step into heating phase and holding time.
Step
Cycles 30x
1
2
3
4
5
6
95.0
95.0
55.0
72.0
72.0
16.0
0:05:00
0:05:00
0:00:30
0:00:30
0:00:30
∞
Step
Step
Step
Step
Step
Temperature [°C]
0
100
20
40
60
ΔT
Δt
80
!
Fig. 23
Graphical display during a program run
The graphical view illustrates the following program parameters:
Program parameters
Display
(example)
Description
Step
Step
1
There is at least one block temperature and
holding time for each program step. All other
parameters are optional.
For more than 6 program steps, use the ar-
row keys to navigate through the displayed
program.
Block Temp. (°C)
Hold Time (h:mm:ss)
95.0
0:05:00
After the heating phase, the sample block
will keep the block temperature for the speci-
fied holding time.
Loops
Cycles 30x
The selected steps are repeated cyclically x
times.
Temperature increment
ΔT
The block temperature is increased or de-
creased by the temperature increment in
each cycle.