Biometra TOne
Operation
41
Annealing Temp.:
Column
Grad. incr.:
Lin.
Std.
Col.
Step
Gradient
3
6
Fig. 19
Linear Gradient Tool (mono block)
}
Linear Gradient Tool: Enter the required annealing temperature in the field
Anneal-
ing Temp.:
.
– The annealing temperature that was entered is applied to a specific column of
the sample block. This is indicated below
Annealing Temp.:
. For the sample block
in the 96 block format, this is
Column
6
.
}
Define a temperature increment under
Grad. incr.:
.
ü
The software will calculate the effective temperatures and display them in a bar
chart.
Lin.
Std.
Col.
Step
Gradient
3
First Column:
Last Column:
Fig. 20
Standard gradient programming
}
Standard gradient programming: Enter the sample block temperature for the first
and the last column in the respective fields.
ü
The software will calculate all effective temperatures and display them in a bar
chart.
}
Confirm your entry by tapping the green check mark.
}
Save the program after entering the data for all program steps.