Specifications
69
Process control
Operation
Stand-alone operation via color touch screen 7"
Remote control application for control and monitoring via
tablet PC
Minimum requirements of the
application regarding the oper-
ating system
iOS 8.0; Android ≥4.4 (API 19)
Minimum requirements of the
application regarding the ther-
mal cycler
Software version V1.43 to V1.44
Interfaces
USB A; Ethernet
Functional scope of the soft-
ware
¡
Generation of Service Info Files (SINF), extended self
test
¡
User-specific quick start of the last programs
¡
Program preview
¡
Multi-step programming with adjustable heating and
cooling rates
¡
Graphical or spreadsheet programming mode
¡
Simplified program creation using protocol templates
¡
Incubation mode
¡
Export function (via USB port)
Time increment
1 to 240 s per cycle
Temperature increment/
decrement
±0.1 to 20 °C per cycle
Program storage capacity
350 programs in 90 user directories
Maximum number of steps per
program
30
Maximum number of cycles per
program
999
Maximum number of nested
cycles
2
Language
German, English, Chinese
Auto restart
Automatic restart after power failure
Electrical specifications
Operating voltage
100 V; 115 V; 230 V ±10 %
Frequency
50 to 60 Hz
Fuses
2 x 8 A H (250 V)
Power consumption, max.
550 VA
Overvoltage category
II
Degree of contamination
2
IP rating of the housing
IP 20
Ambient conditions
Temperature during operation
15 to 35 °C
Humidity during operation
≤70 %
Maximum altitude
2000 m above sea level