ER5000
—
239
Certifications and Warranty
CE
Conforms to all known European Union
requirements.
EMC DIRECTIVE 2014/30/EU
IEC/EN 61000-6-2: 2005
IEC/EN 61000-6-4: 2007/A1:2011
FM
Is not certified but conforms to FM
Standards. FM Class 3600-1998 & FM Class
3615-2006
SPECIAL CONDITIONS FOR SAFE USE:
•
Make sure everything is properly
assembled before powering.
•
Do not open when an explosive
atmosphere is present.
•
Installation to be in accordance with the
latest edition of National Electrical Code.
•
The cable and conduit entry devices
shall be certified flame proof types,
suitable for the conditions of use and
correctly installed.
•
Unused conduit entry must be closed
with suitable blanking element.
•
Seal conduit entry within 18 inches.
•
Rated for 20.5 to 28.5 VDC,
340 mAmps.
•
Max working pressure: 110 PSIG
•
-20°C ≤ T
amb
≤ 60°C
•
Thread entry sizes are 1/8" - 27 NPT
and 1/2" - 14 NPT. Reference page 18
for locations.
•
The user/installer shall not remove any
of the three, 1/8”-27 NPT plugs that
are fitted into the base of the enclosure
as this will invalidate the Hazardous
Location Certification. Reference
page 18 for locations.
Hazardous Locations Special Requirements and Certification for the ER5050
Copies of all certification documents are available at www.tescom.com. Click on the
CERTIFICATIONS
Quick Link.
Summary of Contents for Tescom ER5000 Series
Page 2: ...2 Safety Installation Operations Precautions ...
Page 9: ...9 Table of Contents ...
Page 14: ...14 Features and Specifications ...
Page 25: ...25 What s New ...
Page 29: ...29 How It Works ...
Page 44: ...44 Glossary of Terms ...
Page 57: ...57 Getting Started ...
Page 76: ...76 Installation Variations ...
Page 111: ...111 The ERTune Program ...
Page 116: ...116 The ERTune Program Basic Features ...
Page 187: ...187 The ERTune Program Power User ...
Page 196: ...196 ER5000 Software Development Support ...
Page 202: ...202 Troubleshooting ...
Page 221: ...221 Internal Variables ...
Page 236: ...236 Certifications and Warranty ...
Page 241: ...241 Appendix A Setting up the ERTune Program on Windows 8 PCs ...