Table 5-3: Fuse Requirements
(continued)
Fuse
Function
Rating
Schurter part number
2
Neutral line 110/230
Vac (following mains
filter)
3.15 A, 250 V, fast
acting ceramic
0001.1009
3
Analyzer 12 Vdc supply
rail
5 A, 250 V, fast acting
ceramic
0001.1011
4
Analyzer 24 Vdc supply
rail
3.15 A, 250 V, fast
acting ceramic
0001.9011
5.4.3
Connecting the sample supply and return line
Sample gas supply and sample return connections are
¼
-in. (6 mm) ”Hamlet”
compression tube fittings. To avoid the risk of gas leaks, confirm that these connections
are made correctly and tightly. Both the sample gas supply pipe and the sample return
pipe should be thermal insulated.
The maximum sample gas supply pressure is 2 BarG.
5.4.4
Connecting the signal cables
The signal cables are connected to the system through an entry point outlet C, as shown in
. All signal cables are to be minimum 20 AWG tri-rated switchgear cable.
Customer supplied conduit and cables are to be terminated in the entry point provided in
accordance with local electrical codes.
Reference Manual
Install
00809-0100-4512
October 2019
Reference Manual
61
Summary of Contents for Rosemount CT5100
Page 16: ...16 ...
Page 20: ...Contents Reference Manual October 2019 00809 0100 4512 xx Rosemount CT5100 Ex ...
Page 28: ...Theory of operation Reference Manual October 2019 00809 0100 4512 28 Rosemount CT5100 Ex ...
Page 46: ...Specifications Reference Manual October 2019 00809 0100 4512 46 Rosemount CT5100 Ex ...
Page 66: ...Install Reference Manual October 2019 00809 0100 4512 66 Rosemount CT5100 Ex ...
Page 84: ...Start up procedure Reference Manual October 2019 00809 0100 4512 84 Rosemount CT5100 Ex ...
Page 92: ...Shutdown procedure Reference Manual October 2019 00809 0100 4512 92 Rosemount CT5100 Ex ...
Page 112: ...Gas calibration Reference Manual October 2019 00809 0100 4512 112 Rosemount CT5100 Ex ...