The Rosemount CT5100(Ex) is supplied by Emerson. The gas handling system may be
provided by either you or Emerson, depending upon the specific installation. The circuit
breakers used to control the application of electrical power to the analyzer, the
interconnecting wires, and gas piping are provided by you.
, the items supplied by Emerson are colored blue; the items supplied by you
are colored purple. The green gas handling system may be provided by Emerson or you.
lists the main items of the system.
Figure 3-2: Complete Rosemount CT5100(Ex) Installation
A. Gas handling system
B. Instrument air
C. Sample supply line
D. Sample return (exhaust) line
E. Rosemount CT5100(Ex)
F. Electrical power
G. Two pole main isolator
H. Purge power
I. Control center
J. Measurement data
The analyzer contains an optical system with multiple lasers and a series of optical
components that provide an optical path, a heated multi-pass analysis cell, and sample
and outlet ports that can be connected to a gas handling system and control and analysis
electronics. The number of lasers installed depends upon customer requirements. The
complete system operates from 110 to 230 Vac 5050/60 Hz supply.
The analyzer uses mid-infrared optical absorption spectroscopy to measure gas
concentrations. The light sources are lasers, which are operated to produce wavelength
Description
Reference Manual
October 2019
00809-0100-4512
30
Rosemount CT5100(Ex)
Summary of Contents for Rosemount CT5100
Page 16: ...16 ...
Page 20: ...Contents Reference Manual October 2019 00809 0100 4512 xx Rosemount CT5100 Ex ...
Page 28: ...Theory of operation Reference Manual October 2019 00809 0100 4512 28 Rosemount CT5100 Ex ...
Page 46: ...Specifications Reference Manual October 2019 00809 0100 4512 46 Rosemount CT5100 Ex ...
Page 66: ...Install Reference Manual October 2019 00809 0100 4512 66 Rosemount CT5100 Ex ...
Page 84: ...Start up procedure Reference Manual October 2019 00809 0100 4512 84 Rosemount CT5100 Ex ...
Page 92: ...Shutdown procedure Reference Manual October 2019 00809 0100 4512 92 Rosemount CT5100 Ex ...
Page 112: ...Gas calibration Reference Manual October 2019 00809 0100 4512 112 Rosemount CT5100 Ex ...