Figure 10-9: Input Type
Figure 10-10: Calibration
13. Use the
LEFT, RIGHT, INCR
(increase), and
DECR
(decrease) buttons to enter the
desired offset (in the example above, 1.06 ppm).
14. Press
OK
to confirm.
The factory default span factor is one, and under normal circumstances, the offset
should be between 0.8 and 1.2.
10.2.1
Zero calibrate an individual gas
Procedure
1. To get the
Main menu
, press the
MENU
button on the
Gas Sensor Main
screen
Pressure and Temperature
).
NOTICE
Pressing the
BACK
button will cancel the procedure and return the user to the
previous screen.
Gas calibration
Reference Manual
October 2019
00809-0100-4512
98
Rosemount CT5100(Ex)