NOTICE
Pressing the
BACK
button will cancel the procedure and return the you the previous
screen.
2. Use the middle buttons ( and ) to scroll up or down to select
GAS SERVICE
. Press
the
SEL
button.
Figure 10-30: Main Menu (GAS SERVICE)
3. Use the middle buttons ( and ) to scroll up or down to select
CALIBR PARAMS
or
VERIFY PARAMS
for validation settings). Press the
SEL
button.
Figure 10-31: Gas Service (CALIBR PARAMS)
4. This section applies to systems configured to perform automatic calibration at
timed intervals only.
5. To adjust the frequency of automatic calibration: Use the middle buttons ( and )
to scroll up or down
6. Select
AUTO PERIOD
.
7. Press the
SEL
button.
Reference Manual
Gas calibration
00809-0100-4512
October 2019
Reference Manual
109