Figure 7-5: Display Controller
CAUTION
DO NOT, under any circumstances, press the top two buttons on the display controller
simultaneously. This will override the system code supplied with the analyzer and display
the default program screen for the controller.
This will lock the software controlling the analyzer requiring the analyzer to shut down
externally and restarted risking a possible corruption of the software and loss of the
analyzer operation during the shutdown and restart process.
It is essential that only trained, qualified personnel operate the controls on the analyzer.
Failure to follow these instructions may cause warranty invalidation, property damage
and/or personal injury or death.
Figure 7-6: Default Program Screen
) is used to program the operation of the overpressure
air system that prevents flammable and explosive gases from entering the electrical
compartment of the analyzer. Adjustment of the overpressure air system must only be
performed by maintenance personnel.
Start-up procedure
Reference Manual
October 2019
00809-0100-4512
82
Rosemount CT5100(Ex)
Summary of Contents for Rosemount CT5100
Page 16: ...16 ...
Page 20: ...Contents Reference Manual October 2019 00809 0100 4512 xx Rosemount CT5100 Ex ...
Page 28: ...Theory of operation Reference Manual October 2019 00809 0100 4512 28 Rosemount CT5100 Ex ...
Page 46: ...Specifications Reference Manual October 2019 00809 0100 4512 46 Rosemount CT5100 Ex ...
Page 66: ...Install Reference Manual October 2019 00809 0100 4512 66 Rosemount CT5100 Ex ...
Page 84: ...Start up procedure Reference Manual October 2019 00809 0100 4512 84 Rosemount CT5100 Ex ...
Page 92: ...Shutdown procedure Reference Manual October 2019 00809 0100 4512 92 Rosemount CT5100 Ex ...
Page 112: ...Gas calibration Reference Manual October 2019 00809 0100 4512 112 Rosemount CT5100 Ex ...