11.8
Visual examination
If the BIST fails to identify the problem, the next step is a visual examination of the
Rosemount CT5100(Ex).
Procedure
1. Open the door of the electrical compartment and remove the top cover.
.
2. Visually examine the exterior of the analyzer for signs of damage.
3. Perform a visual inspection of the optical and electrical components inside the
analyzer.
Refer to
.
4. If any loose connections are found in the electrical compartment, refer to the wiring
) to identify and repair the connection.
5. Refit the top cover and close the electrical compartment door.
11.9
Functional failure diagnostics
If the Built-in-Self Test and the visual examination fail to identify the fault, perform the
failure diagnostics and recommended actions.
11.9.1
No illuminated LEDs on Power supply unit (PSU) when
analyzer switched ON.
Recommended actions
1. Check mains power is available.
2. If OK, replace PSUs.
11.9.2
Nothing powers up when connected to mains
Recommended actions
1. Check AC fuses (number 1 and number 2)
2. Replace if necessary.
11.9.3
Only MOXA and BEKA powers up
Condition
The startup screen displays BEKA; all other electronic modules remain unpowered.
Recommended actions
1. Check the 12 V PSU is working (LED illuminated).
2. If the 12 V PSU is not working, replace it.
Troubleshooting and diagnostics
Reference Manual
October 2019
00809-0100-4512
122
Rosemount CT5100(Ex)
Summary of Contents for Rosemount CT5100
Page 16: ...16 ...
Page 20: ...Contents Reference Manual October 2019 00809 0100 4512 xx Rosemount CT5100 Ex ...
Page 28: ...Theory of operation Reference Manual October 2019 00809 0100 4512 28 Rosemount CT5100 Ex ...
Page 46: ...Specifications Reference Manual October 2019 00809 0100 4512 46 Rosemount CT5100 Ex ...
Page 66: ...Install Reference Manual October 2019 00809 0100 4512 66 Rosemount CT5100 Ex ...
Page 84: ...Start up procedure Reference Manual October 2019 00809 0100 4512 84 Rosemount CT5100 Ex ...
Page 92: ...Shutdown procedure Reference Manual October 2019 00809 0100 4512 92 Rosemount CT5100 Ex ...
Page 112: ...Gas calibration Reference Manual October 2019 00809 0100 4512 112 Rosemount CT5100 Ex ...