CAUTION
EQUIPMENT DAMAGE
When preparing the analyzer for transport by air, road, or rail, safeguard the analyzer
against movement or break-away during transport by securely strapping it in place.
Failure to observe may cause damage to the equipment.
WARNING
EXPLOSION HAZARD
In accordance with IEC 60079-0: 2011 Clause 30.
Installing and wiring the analyzer must comply with all relevant national legislative
requirements and regulations.
Consider all safety instructions within this manual and all associated analyzer instruction
manuals.
WARNING
EXPLOSION HAZARD
In accordance with IEC 60079-0: 2011 Clause 30.
Installing the analyzer requires opening the enclosure and working at the open unit. This is
permitted only when both the analyzer and connected external circuitry are de-energized.
Depending on the local regulation, this may require a competent hot work supervisor to
issue a hot work permit.
WARNING
HEAVY ITEM
In accordance with IEC 60079-0: 2011 Clause 30.
Failure to propery handle the analyzer may cause injury to personnel.
Ensure the wall the analyzer is mounted on is solid, stable, and of suitable material to
hold the weight of the analyzer.
Handle the analyzer with caution during unpacking, installing, maintaining, and
transporting to prevent crushing of hands, feet, or other body parts.
The analyzer weighs 121 lb. (55 kg).
Emerson recommends that a minimum of two people move and lift the analyzer.
Wear suitable protective gloves and protective footwear.
Reference Manual
Install
00809-0100-4512
October 2019
Reference Manual
49
Summary of Contents for Rosemount CT5100
Page 16: ...16 ...
Page 20: ...Contents Reference Manual October 2019 00809 0100 4512 xx Rosemount CT5100 Ex ...
Page 28: ...Theory of operation Reference Manual October 2019 00809 0100 4512 28 Rosemount CT5100 Ex ...
Page 46: ...Specifications Reference Manual October 2019 00809 0100 4512 46 Rosemount CT5100 Ex ...
Page 66: ...Install Reference Manual October 2019 00809 0100 4512 66 Rosemount CT5100 Ex ...
Page 84: ...Start up procedure Reference Manual October 2019 00809 0100 4512 84 Rosemount CT5100 Ex ...
Page 92: ...Shutdown procedure Reference Manual October 2019 00809 0100 4512 92 Rosemount CT5100 Ex ...
Page 112: ...Gas calibration Reference Manual October 2019 00809 0100 4512 112 Rosemount CT5100 Ex ...