4. The system will work through the
ZERO
and
SPAN
calibration process for all
readings.
The
Please wait
screen will be displayed. This takes approximately one minute per
gas depending on the purge time.
Figure 10-28: Calibration
The
Finished
screen displays when the Zero and Span calibrations process
completes.
Figure 10-29: Calibration (Finished)
5. Press
BACK
.
This completes the process, and the display returns to the
Main menu
.
10.2.4
Change the automatic calibration settings
Procedure
1. To get to the
Main menu
, press the
MENU
button on the
Gas Sensor Main
screen
Pressure and Temperature
screen (
).
Gas calibration
Reference Manual
October 2019
00809-0100-4512
108
Rosemount CT5100(Ex)