Figure 5-6: Power Gland
A. Purge power and alarm entry point
B. Power entry point
Two sources of electrical power are applied to the analyzer:
•
A purge electrical and alarm supply that is used to power the overpressure facility in the
electrical compartment of the analyzer. Refer to
•
An instrumentation electrical supply that is used to power all other functions of the
analyzer. See
The customer supplied circuit breakers
must
be in accordance with ATEX / IECEx / North
American protection concepts. The main power isolator controls the application of
electrical power to the purge controller with the secondary power isolator controlling the
application of electrical power to the remaining analyzer functions.