4. Disconnect the wiring harness (H).
5. Remove and retain the four screws (B) nuts (G), plain washers (D) and spring
washers (F).
6. Remove and discard the unserviceable detector module (C).
7. Inspect the new detector module for damage.
8. Fit the new detector module in position on the detector bracket (E). Secure the
detector module with the four screws, nuts and washers retained in
G).
9. Connect the wiring harness (H) to the detector module (C).
10. Connect the trigger OUT harness (A) to the detector module (C).
11. Close the door of the electrical compartment and restart the analyzer.
12.15
Replacing the O
2
detector
The O
2
detector is located on the underside of the optical compartment base plate. First
determine which detector type is fitted: 1 or 2. Replace the O
2
detector identified using
the following steps.
12.15.1
Replacing O
2
detector (type 1)
Procedure
1. Shut down the analyzer. Refer to
2. Open the door of the electrical compartment.
3. Disconnect the Bayonet Neill–Concelman (BNC) cable and the Power cable from the
O
2
detector (D).
Reference Manual
Repair and replacement
00809-0100-4512
October 2019
Reference Manual
147
Summary of Contents for Rosemount CT5100
Page 16: ...16 ...
Page 20: ...Contents Reference Manual October 2019 00809 0100 4512 xx Rosemount CT5100 Ex ...
Page 28: ...Theory of operation Reference Manual October 2019 00809 0100 4512 28 Rosemount CT5100 Ex ...
Page 46: ...Specifications Reference Manual October 2019 00809 0100 4512 46 Rosemount CT5100 Ex ...
Page 66: ...Install Reference Manual October 2019 00809 0100 4512 66 Rosemount CT5100 Ex ...
Page 84: ...Start up procedure Reference Manual October 2019 00809 0100 4512 84 Rosemount CT5100 Ex ...
Page 92: ...Shutdown procedure Reference Manual October 2019 00809 0100 4512 92 Rosemount CT5100 Ex ...
Page 112: ...Gas calibration Reference Manual October 2019 00809 0100 4512 112 Rosemount CT5100 Ex ...