K. Ribbon cable (to peripheral PCB)
L. 12 V power out connector J17
M. 12 V power in connector J16
N. 12 V power input wiring harness
O. 12 V power output wiring harness
P. Laser module ribbon cable
Q. HMI Display wiring harness
R. Connector J22
S. Ethernet cable
T. Ethernet connector
Procedure
1. Shut down the analyzer.
.
2. Open the door of the electrical compartment.
3. Tag or otherwise identify all wiring harnesses before disconnecting.
4. On the motherboard, disconnect the TEC board ribbon cable (E) from connector J12
(B).
Refer to
5. Disconnect the Detector Module 1 Signal Out wiring harness (I) from connector J31
(G).
NOTICE
The Detector Module 2 Signal Out wiring harness (H) is an optional item that only
forms part of those analyzers that are fitted with an O
2
detector.
6. Disconnect the Detector Module 2 Signal Out wiring harness (H) from connector
J32 (F).
7. Disconnect the Laser Module ribbon cables (P) from the motherboard.
The number of Laser Module ribbon cables to be disconnected will vary depending
upon the application that the analyzer has been configured for.
8. Disconnect the 12 V power output wiring harness (O) from connector J17 (L).
9. Disconnect the 12 V power input wiring harness (N) from connector J16 (M).
10. Disconnect the peripheral PCB ribbon cable (K) from connector J7 (J).
11. Disconnect the Ethernet cable (S) from the Ethernet connector (T).
12. Disconnect the HMI display wiring harness (Q) from connector J22 (R).
13. Remove and retain the six screws (D) and associated spring washers (C) and flat
washers (A).
14. Remove and discard the unserviceable motherboard.
15. Inspect the replacement motherboard for signs of damage or delamination.
16. Place the motherboard in position on the backplate and secure by fitting the six
screws (D) and the associated spring washers (C) and flat washers (A) retained in
. Torque tighten the screws to 0.6 Nm.
Repair and replacement
Reference Manual
October 2019
00809-0100-4512
138
Rosemount CT5100(Ex)
Summary of Contents for Rosemount CT5100
Page 16: ...16 ...
Page 20: ...Contents Reference Manual October 2019 00809 0100 4512 xx Rosemount CT5100 Ex ...
Page 28: ...Theory of operation Reference Manual October 2019 00809 0100 4512 28 Rosemount CT5100 Ex ...
Page 46: ...Specifications Reference Manual October 2019 00809 0100 4512 46 Rosemount CT5100 Ex ...
Page 66: ...Install Reference Manual October 2019 00809 0100 4512 66 Rosemount CT5100 Ex ...
Page 84: ...Start up procedure Reference Manual October 2019 00809 0100 4512 84 Rosemount CT5100 Ex ...
Page 92: ...Shutdown procedure Reference Manual October 2019 00809 0100 4512 92 Rosemount CT5100 Ex ...
Page 112: ...Gas calibration Reference Manual October 2019 00809 0100 4512 112 Rosemount CT5100 Ex ...