11.5
Tools and test equipment
All tools required when performing failure diagnosis are standard hand tools.
The only test equipment required when performing failure diagnosis is a multi-meter that
may be used to perform continuity checks on electrical wiring during an inspection.
11.6
General troubleshooting and diagnostics
information
The Rosemount
™
CT5100(Ex) is specifically designed to run unattended indefinitely, to
automatically resolve system issues, and to recover from power failures and return to a
normal working state. This troubleshooting guide is intended to assist maintenance
personnel when the analyzer has not appeared to be working normally for a period of
more than five minutes. If the procedures given in this section fail to return the analyzer to
normal operation or do not identify a fault, notify your service agent for further assistance.
11.7
Diagnostics menu
Selecting
DIAGNOSTICS
on the
Main Menu
will open the
Diagnostics Menu
).
The
Diagnostics Menu
lists those main components of the Rosemount CT5100(Ex) where
problems can be diagnosed using the BIST system. The
Diagnostics Menu
also enables you
to check on the status and, where appropriate, the values of, any of the listed
components.
Press the
REFR
button (F) to refresh the fault list.
Use the
Scroll Up
(C) and
Scroll Down
(G) buttons to highlight a component for further
investigation by the BIST system.
Once you have highlighted the component you wish to diagnose, press the
SEL
button (B)
to select that fault. The
Diagnostics
screen for that component will then open.
Reference Manual
Troubleshooting and diagnostics
00809-0100-4512
October 2019
Reference Manual
119
Summary of Contents for Rosemount CT5100
Page 16: ...16 ...
Page 20: ...Contents Reference Manual October 2019 00809 0100 4512 xx Rosemount CT5100 Ex ...
Page 28: ...Theory of operation Reference Manual October 2019 00809 0100 4512 28 Rosemount CT5100 Ex ...
Page 46: ...Specifications Reference Manual October 2019 00809 0100 4512 46 Rosemount CT5100 Ex ...
Page 66: ...Install Reference Manual October 2019 00809 0100 4512 66 Rosemount CT5100 Ex ...
Page 84: ...Start up procedure Reference Manual October 2019 00809 0100 4512 84 Rosemount CT5100 Ex ...
Page 92: ...Shutdown procedure Reference Manual October 2019 00809 0100 4512 92 Rosemount CT5100 Ex ...
Page 112: ...Gas calibration Reference Manual October 2019 00809 0100 4512 112 Rosemount CT5100 Ex ...