12.21
Cleaning the cell mirrors
This procedure should be used during service intervention and only attempted if there has
been a decrease in the laser pulse amplitudes which is likely to have been caused by
contamination of the cell.
CAUTION
Anti-static precautions in the form of a static dissipative mat and anti-static wrist strap
must be used at all times when handling electronic components and assemblies.
12.21.1
Equipment materials
The following equipment and materials are required when cleaning the cell mirrors:
•
Spectroscopic grade methanol or ethanol
•
Optical grade lens tissue
•
Invertible air duster
•
Hex head keys, sizes 1.5 mm to 4 mm
Repair and replacement
Reference Manual
October 2019
00809-0100-4512
156
Rosemount CT5100(Ex)