Set-up totalizing on the Rosemount 3492
Menu
SETUP
→
[CONTROLLER]
→
OUTPUT
→
TOTALIZER
and
MONITOR
→
[CONTROLLER]
→
READINGS
→
TOTALIZER
The Rosemount 3492 has two independent, internal 8-digit totalizers, Totalizer 1 and
Totalizer 2, which are updated several times every second.
Set-up Totalizer 1
Note
The totalizer operates with an input of control unit PV (
D800
) in units per second. If the PV
is a volumetric flow rate (e.g. m
3
/hour), the totalizer can accumulate this flow volume and
give the total volume throughput.
Table 4-33: Totalizer 1 Parameters
Parameter
Description
Factory default
P530
Total 1 factor
One count is added to the Totalizer count (D828) for
a quantity of flow as defined by this parameter. The
unit of measurement for D828 is defined by
parameter P531.
0.0
P531
Total 1 units
This parameter defines the units for the Totalizer 1
Count (D828).
When using totalizer units of gal, an 8-digit totalizer
increments rapidly and rolls over too frequently. To
help, alternative totalizer units of galx10, galx100,
galx1000, and Mgal are available for selection. When
these special units are selected after selecting gal
units, the totalizer factor (P530) is automatically re-
scaled by x10, x100, x1000, or x1000000 depending
on the selection. Similarly, re-scaling is automatic for
ltrx10, ltrx100, etc. after selecting ltr as totalizer
units.
See
Using the App Wizard to set-up a flow
for a flow application example where gal
x 100 totalizer units are needed.
None
P534
Pulse Width
A Totalizer mode relay is energized for a
programmed duration (P534) each time the
Totalizer count (D828 or D829) is incremented.
P534 controls the pulse width (i.e. the pulse 'on'
time) and is between 10 ms and 2.5 s, changeable in
steps of 10 ms.
Parameter P534 also defines the pulse width for a
Sampler mode relay (see
).
100 ms
D828
Totalizer 1
This read-only parameter displays the Totalizer 1
Count. To add this to the
Full PV
display, see
N/A
Getting started
Reference Manual
January 2021
00809-0100-4841
126
Emerson.com/Rosemount
Summary of Contents for Rosemount 3490 Series
Page 4: ...4 ...
Page 8: ...Introduction Reference Manual January 2021 00809 0100 4841 8 Emerson com Rosemount ...
Page 16: ...Control unit overview Reference Manual January 2021 00809 0100 4841 16 Emerson com Rosemount ...
Page 36: ...Installation Reference Manual January 2021 00809 0100 4841 36 Emerson com Rosemount ...
Page 134: ...Getting started Reference Manual January 2021 00809 0100 4841 134 Emerson com Rosemount ...
Page 184: ...Additional features Reference Manual January 2021 00809 0100 4841 184 Emerson com Rosemount ...
Page 189: ...Reference Manual 00809 0100 4841 January 2021 Emerson com Rosemount 189 ...