Set-up Totalizer 2 (daily totalized flow)
This is an 8-digit totalizer showing daily total flow (midnight to midnight) through a flow
structure. It is automatically re-set to zero at each midnight.
Note
A control unit relay can be configured to output a pulse for each increment to the Totalizer
Count parameter. See
for how to do this.
Table 4-36: Totalizer 2 Parameters
Parameter
Description
Factory default
P532
Total 2 Factor
One count is added to the Totalizer 2 Count (D829)
for a quantity of liquid as defined by this parameter.
The display unit of measurement for D829 is defined
by P533.
0.0
P533
Total 2 Units
This parameter defines the units for the Totalizer 2
Count (D829).
None
P534
Pulse Width
A Totalizer mode relay is energized for a
programmed duration (P534) each time the
Totalizer Count (D828 or D829) is incremented.
P534 controls the pulse width (i.e. the pulse 'on'
time) and is between 10 ms and 2.5 s, changeable in
steps of 10 ms.
Parameter P534 also defines the pulse width for a
Sampler mode relay (see
).
Note
mode relays.
100 ms
P536
Total 2 Source
Use this to select whether the control unit PV
(D800), SV (D801), TV (D802) or FV (D803) value is
to be totalized over a 24 hours period for the
Totalizer 2 Count (D829). Alternatively, select None
if not requiring Totalizer 2.
See
Optional change: transmitter input channel
and FV.
MCU PV
P537
Total 2 dec pl
Use this to set the number of decimal places to be
shown when displaying the Totalizer 2 Count (D829).
1
D829
Totalizer 2
This read-only parameter displays the Totalizer 2
count, which is the daily flow total for the present
day. It is reset to zero at midnight.
To add this value to the PV display, see
display line, use the
left arrow
or
right arrow
buttons to scroll through daily flow totals from the
previous 365 days. The
Esc
button restores the upper
display line to show the daily flow total value for the
present day.
Reference Manual
Getting started
00809-0100-4841
January 2021
Emerson.com/Rosemount
129
Summary of Contents for Rosemount 3490 Series
Page 4: ...4 ...
Page 8: ...Introduction Reference Manual January 2021 00809 0100 4841 8 Emerson com Rosemount ...
Page 16: ...Control unit overview Reference Manual January 2021 00809 0100 4841 16 Emerson com Rosemount ...
Page 36: ...Installation Reference Manual January 2021 00809 0100 4841 36 Emerson com Rosemount ...
Page 134: ...Getting started Reference Manual January 2021 00809 0100 4841 134 Emerson com Rosemount ...
Page 184: ...Additional features Reference Manual January 2021 00809 0100 4841 184 Emerson com Rosemount ...
Page 189: ...Reference Manual 00809 0100 4841 January 2021 Emerson com Rosemount 189 ...