![Emerson Rosemount 3490 Series Reference Manual Download Page 81](http://html.mh-extra.com/html/emerson/rosemount-3490-series/rosemount-3490-series_reference-manual_2400788081.webp)
to m
3
/hr and the display configured to show the SV (level in cm on top line) and totalizer in
m
3
(on bottom line). The transmitter base units must be set to meters.
Table 4-3: Pre-programmed Flow Calculation Factors (Parshall Flumes)
Profile (P113 or P123)
Pwr (P115 or P125)
k (P115 or P125)
mul (P115 or P125)
P116 (or P126)
Parshall 1-in
1.550
151.7
1.0
1.0
Parshall 2-in
1.550
303.4
1.0
1.0
Parshall 3-in
1.547
445.2
1.0
1.0
Parshall 6-in
1.580
924.5
1.0
1.0
Parshall 9-in
1.530
1378
1.0
1.0
Parshall 1 ft
1.522
1795
1.0
1.0
Parshall 1.5 ft
1.538
2693
1.0
1.0
Parshall 2 ft
1.550
3590
1.0
1.0
Parshall 3 ft
1.566
5386
1.0
1.0
Parshall 4 ft
1.578
7181
1.0
1.0
Parshall 5 ft
1.587
8976
1.0
1.0
Parshall 6 ft
1.595
10770
1.0
1.0
Parshall 8 ft
1.607
14360
1.0
1.0
Parshall 10 ft
1.600
17672
1.0
1.0
Parshall 12 ft
1.600
20982
1.0
1.0
Related information
Optional change: transmitter input channel settings (advanced users)
Using a plotted profile for calculating volume or flow
When parameter P113 (or P123) is set to “special”, the parameter P115 (or P125) is used
for defining a 20-point look-up table that represents the profile of a vessel or open flow
channel that is not in the control unit library.
Note
When using the App Wizard, select
20 point look-up
at the prompt for a profile.
Each point is a Cartesian co-ordinate (X, Y). The X value represents a level, and the Y value
is the corresponding volume or flow rate. The X points are at user-defined intervals,
typically in equal increments (5%) of maximum height.
These X points may be entered as levels (in the same units as the transmitter) and the Y
points entered as the corresponding volumes or flow rates for those levels (in the same
units as selected for the control unit Primary/Process Value). See
, right-hand
graph.
Reference Manual
Getting started
00809-0100-4841
January 2021
Emerson.com/Rosemount
81
Summary of Contents for Rosemount 3490 Series
Page 4: ...4 ...
Page 8: ...Introduction Reference Manual January 2021 00809 0100 4841 8 Emerson com Rosemount ...
Page 16: ...Control unit overview Reference Manual January 2021 00809 0100 4841 16 Emerson com Rosemount ...
Page 36: ...Installation Reference Manual January 2021 00809 0100 4841 36 Emerson com Rosemount ...
Page 134: ...Getting started Reference Manual January 2021 00809 0100 4841 134 Emerson com Rosemount ...
Page 184: ...Additional features Reference Manual January 2021 00809 0100 4841 184 Emerson com Rosemount ...
Page 189: ...Reference Manual 00809 0100 4841 January 2021 Emerson com Rosemount 189 ...