![Emerson Rosemount 3490 Series Reference Manual Download Page 83](http://html.mh-extra.com/html/emerson/rosemount-3490-series/rosemount-3490-series_reference-manual_2400788083.webp)
Example
The Rosemount 3491 screens are shown here.
MAIN MENU
Run App?
MONITOR
SETUP
SELECT INSTRUMENT
Tx1 : ****
#1
Controller:
****
SETUP
INPUT CHANNEL
INPUT CHANNEL
Ch1 Profile
Ch1 Pre Scale
Ch1 NLP Data
Ch1 NLP Data P115
x0: 0.000
y0: 0.000
DISPLAY
OUTPUT
x3
x4
The
SELECT INSTRUMENT
menu does not appear unless a HART transmitter is
connected.
2. Press the
red (enter)
button once to select point 0.
Before editing, you can use the
down arrow
button to page down to another point
or use the
Esc
button to return to the menu.
3. With
X0:
highlighted, use the
right arrow
button to move across to the X0 value.
4. Use the
up arrow
or
down arrow
buttons to edit the X0 value.
5. Press the
red (enter)
button once to confirm the new X0 value.
6. With
Y0:
highlighted, use the
right arrow
button to move across to the Y0 value.
7. Use the
up arrow
or
down arrow
buttons to edit the Y0 value.
8. Press the
red (enter)
button once to confirm the new Y0 value.
9. To edit X0 and Y0 again, return to
down arrow
button to
display the page with X1 and Y1 or use the
Esc
button to return to the menu.
10. Repeat for all points in the profile.
An X point with 0.0 terminates the profile (unless it is point X0, Y0).
Need help?
The
Esc
button can be used during editing to abort and restore the original values; it needs
to be selected again to move back to the 'X' or 'Y' text.
4.4.10
Digital inputs IN1 and IN2
Digital inputs IN1 and IN2 can be programmed to perform an action whenever they are
triggered.
Note
Digital input IN1 and IN2 statuses are shown on the left-hand side of the display: o =
inactive or
▸
= active.
Parameter D835 shows the statuses of the inputs: active (1) or inactive (0). First digit
represents IN1. The second digit represents IN2.
Reference Manual
Getting started
00809-0100-4841
January 2021
Emerson.com/Rosemount
83
Summary of Contents for Rosemount 3490 Series
Page 4: ...4 ...
Page 8: ...Introduction Reference Manual January 2021 00809 0100 4841 8 Emerson com Rosemount ...
Page 16: ...Control unit overview Reference Manual January 2021 00809 0100 4841 16 Emerson com Rosemount ...
Page 36: ...Installation Reference Manual January 2021 00809 0100 4841 36 Emerson com Rosemount ...
Page 134: ...Getting started Reference Manual January 2021 00809 0100 4841 134 Emerson com Rosemount ...
Page 184: ...Additional features Reference Manual January 2021 00809 0100 4841 184 Emerson com Rosemount ...
Page 189: ...Reference Manual 00809 0100 4841 January 2021 Emerson com Rosemount 189 ...