Table 4-31: Control Unit Reporting of Faults
Source
Cause
As seen on
screen
Status
LED
Full PV
display
Relay
output
Current
output
Alarm
report
(D830)
Fault
report
(D831)
3490
Rom checksum error
ROM Error
Constant
Yes
Yes
Yes
N/A
Yes
RAM test error
RAM Error
Constant
Yes
Yes
Yes
N/A
Yes
Real Time Clock Fault
Clock fault
Constant
Yes
Yes
Yes
N/A
Yes
EEPROM Signature Error
EEPROM Sig err
Constant
Yes
Yes
Yes
N/A
Yes
EEPROM checksum error
EEPROM CKS err
Constant
Yes
Yes
Yes
N/A
Yes
ADC error
ADC_error
Constant
Yes
Yes
Yes
N/A
Yes
Control Unit temperature
out-of-limits
CU Temp OL
Constant
Yes
Yes
N/A
Yes
Xmtr
Field Device Malfunction
Xmtr Fault
Constant
Yes
Yes
Yes
N/A
Yes
4.4.15
Set-up totalizing
The totalizer can be easily set-up using the Totalizer Wizard, accessible by navigating to
the
TOTALIZER
menu.
The Totalizer Wizard also forms a part of the Application (App) Wizard.
Note
The wizard requires the totalizer source parameter to have suitable units selected e.g.
m
3
/hour. If no suitable units are selected, an 'invalid units' message appears.
Procedure
Navigate to
SETUP
→
CONTROLLER
→
OUTPUT
→
TOTALIZER
→
Totalizer Wizard
.
Example
The Rosemount 3491 screens are shown here.
MAIN MENU
Run App?
MONITOR
SETUP
SELECT INSTRUMENT
Tx1: ****
1
Controller: ****
SETUP
APPLICATION
DISPLAY
OUTPUT
OUTPUT
CURRENT OUTPUT
TOTALIZER
1
TOTALIZER
Total Factor
Total Units
Totalizer Wizard
RELAY
The
SELECT INSTRUMENT
menu does not appear unless a HART transmitter is connected.
After using a wizard, adjustments can be made to the totalizer parameters.
Getting started
Reference Manual
January 2021
00809-0100-4841
124
Emerson.com/Rosemount
Summary of Contents for Rosemount 3490 Series
Page 4: ...4 ...
Page 8: ...Introduction Reference Manual January 2021 00809 0100 4841 8 Emerson com Rosemount ...
Page 16: ...Control unit overview Reference Manual January 2021 00809 0100 4841 16 Emerson com Rosemount ...
Page 36: ...Installation Reference Manual January 2021 00809 0100 4841 36 Emerson com Rosemount ...
Page 134: ...Getting started Reference Manual January 2021 00809 0100 4841 134 Emerson com Rosemount ...
Page 184: ...Additional features Reference Manual January 2021 00809 0100 4841 184 Emerson com Rosemount ...
Page 189: ...Reference Manual 00809 0100 4841 January 2021 Emerson com Rosemount 189 ...