Alert num-
ber
Alert title
Possible cause
Recommended actions
A102
Drive Overrange
The drive power (current/voltage) is
at its maximum.
•
Check the drive gain and the pickoff
voltage.
•
Check the wiring between the sensor
and the transmitter.
•
Verify that internal wiring is secure and
that there are no internal electrical
problems.
•
Check for air in the flow tubes, tubes
not filled, foreign material in the tubes,
coating in the tubes, or other process
problems.
•
Check for fluid separation by monitor-
ing the density value and comparing
the results against expected density val-
ues.
•
Ensure that the sensor orientation is ap-
propriate for your application. Settling
from a two-phase or three-phase fluid
can cause this alert even if the flow
tubes are full.
A104
Calibration in Pro-
gress
A calibration procedure is in proc-
ess.
•
Allow the procedure to complete.
•
For zero calibration, you may abort the
calibration, set
Zero Time
to a lower val-
ue, and restart the calibration.
A105
Two-Phase Flow
The line density is outside the user-
defined two-phase flow limits.
•
Check for two-phase flow.
A107
Power Reset Occur-
red
The transmitter has been restarted.
•
No action required.
•
If desired, you can set
Alert Severity Level
to
Ignore
.
N/A
Density D1 Calibra-
tion in Progress
A D1 density calibration is in pro-
gress.
•
No action required.
N/A
Density D2 Calibra-
tion in Progress
A D2 density calibration is in pro-
gress.
•
No action required.
N/A
Zero Calibration in
Progress
A zero calibration is in progress.
•
No action required.
N/A
Reverse Flow
Flow through the device is in the re-
verse direction (against the flow ar-
row).
•
No action required.
Troubleshooting
74
Micro Motion
®
LNG Series Meters
Summary of Contents for Micro Motion LNG Series
Page 1: ...Configuration and Use Manual MMI 20032809 Rev AA February 2017 Micro Motion LNG Series Meters ...
Page 6: ...Contents iv Micro Motion LNG Series Meters ...
Page 8: ...Getting Started 2 Micro Motion LNG Series Meters ...
Page 12: ...Before you begin 6 Micro Motion LNG Series Meters ...
Page 18: ...Quick start 12 Micro Motion LNG Series Meters ...
Page 20: ...Configuration and commissioning 14 Micro Motion LNG Series Meters ...
Page 50: ...Configure device options and preferences 44 Micro Motion LNG Series Meters ...
Page 56: ...Integrate the meter with the control system 50 Micro Motion LNG Series Meters ...
Page 58: ...Complete the configuration 52 Micro Motion LNG Series Meters ...
Page 60: ...Operations maintenance and troubleshooting 54 Micro Motion LNG Series Meters ...
Page 74: ...Measurement support 68 Micro Motion LNG Series Meters ...
Page 94: ...Using ProLink III with the transmitter 88 Micro Motion LNG Series Meters ...
Page 98: ...Default values and ranges 92 Micro Motion LNG Series Meters ...
Page 99: ...Default values and ranges Configuration and Use Manual 93 ...