To know whether your pickoff voltage is unusually low, you must collect pickoff voltage
data during the problem condition and compare it to pickoff voltage data from a period of
normal operation.
Possible causes and recommended actions for low pickoff voltage
Table 10-7:
Possible cause
Recommended actions
Faulty wiring runs between the sensor
and transmitter
Verify wiring between sensor and transmitter.
Process flow rate beyond the limits of
the sensor
Verify that the process flow rate is not out of range of the
sensor.
The vibrating element is not vibrating
•
Check for plugging or deposition.
•
Ensure that the vibrating element is free to vibrate
(no mechanical binding).
•
Verify wiring.
•
.
Moisture in the sensor electronics
Eliminate the moisture in the sensor electronics.
The sensor is damaged, or sensor mag-
nets may have become demagnetized
Replace the sensor.
10.11.1
Collect pickoff voltage data
Pickoff voltage data can be used to diagnose a variety of process and equipment
conditions. Collect pickoff voltage data from a period of normal operation, and use this
data as a baseline for troubleshooting.
Procedure
1.
Navigate to the pickoff voltage data.
2.
Observe and record data for both the left pickoff and the right pickoff, over an
appropriate period of time, under a variety of process conditions.
10.12
Check for internal electrical problems
Shorts between sensor terminals or between the sensor terminals and the sensor case can
cause the sensor to stop working.
Possible causes and recommended actions for electrical shorts
Table 10-8:
Possible cause
Recommended action
Moisture inside the sensor junction
box
Ensure that the junction box is dry and no corrosion is
present.
Liquid or moisture inside the sensor
case
Contact Micro Motion.
Troubleshooting
82
Micro Motion
®
LNG Series Meters
Summary of Contents for Micro Motion LNG Series
Page 1: ...Configuration and Use Manual MMI 20032809 Rev AA February 2017 Micro Motion LNG Series Meters ...
Page 6: ...Contents iv Micro Motion LNG Series Meters ...
Page 8: ...Getting Started 2 Micro Motion LNG Series Meters ...
Page 12: ...Before you begin 6 Micro Motion LNG Series Meters ...
Page 18: ...Quick start 12 Micro Motion LNG Series Meters ...
Page 20: ...Configuration and commissioning 14 Micro Motion LNG Series Meters ...
Page 50: ...Configure device options and preferences 44 Micro Motion LNG Series Meters ...
Page 56: ...Integrate the meter with the control system 50 Micro Motion LNG Series Meters ...
Page 58: ...Complete the configuration 52 Micro Motion LNG Series Meters ...
Page 60: ...Operations maintenance and troubleshooting 54 Micro Motion LNG Series Meters ...
Page 74: ...Measurement support 68 Micro Motion LNG Series Meters ...
Page 94: ...Using ProLink III with the transmitter 88 Micro Motion LNG Series Meters ...
Page 98: ...Default values and ranges 92 Micro Motion LNG Series Meters ...
Page 99: ...Default values and ranges Configuration and Use Manual 93 ...