Standard Gas Density
•
Gas Standard Volume Flow Measurement Unit
•
Gas Standard Volume Flow Cutoff
Restriction
You cannot implement both liquid volume flow and gas standard volume flow at the same time. You
must choose one or the other.
4.3.1
Configure
Volume Flow Type
for gas applications
ProLink III
Device Tools
>
Configuration
>
Process Measurement
>
Flow
Overview
Volume Flow Type
controls whether liquid or gas standard volume flow measurement is
used.
Procedure
Set
Volume Flow Type
to
Gas Standard Volume
.
4.3.2
Configure
Standard Gas Density
ProLink III
Device Tools
>
Configuration
>
Process Measurement
>
Flow
Overview
The
Standard Gas Density
value is used to convert the measured flow data to the standard
reference values.
Prerequisites
Ensure that
Density Measurement Unit
is set to the measurement unit you want to use for
Standard Gas Density
.
Procedure
Set
Standard Gas Density
to the standard reference density of the gas you are measuring.
Note
ProLink III provides a guided method that you can use to calculate the standard density of your gas, if
you do not know it.
Configure process measurement
Configuration and Use Manual
25
Summary of Contents for Micro Motion LNG Series
Page 1: ...Configuration and Use Manual MMI 20032809 Rev AA February 2017 Micro Motion LNG Series Meters ...
Page 6: ...Contents iv Micro Motion LNG Series Meters ...
Page 8: ...Getting Started 2 Micro Motion LNG Series Meters ...
Page 12: ...Before you begin 6 Micro Motion LNG Series Meters ...
Page 18: ...Quick start 12 Micro Motion LNG Series Meters ...
Page 20: ...Configuration and commissioning 14 Micro Motion LNG Series Meters ...
Page 50: ...Configure device options and preferences 44 Micro Motion LNG Series Meters ...
Page 56: ...Integrate the meter with the control system 50 Micro Motion LNG Series Meters ...
Page 58: ...Complete the configuration 52 Micro Motion LNG Series Meters ...
Page 60: ...Operations maintenance and troubleshooting 54 Micro Motion LNG Series Meters ...
Page 74: ...Measurement support 68 Micro Motion LNG Series Meters ...
Page 94: ...Using ProLink III with the transmitter 88 Micro Motion LNG Series Meters ...
Page 98: ...Default values and ranges 92 Micro Motion LNG Series Meters ...
Page 99: ...Default values and ranges Configuration and Use Manual 93 ...