Do not terminate the shielding inside the wiring compartment.
-
360-degree termination of shielding is unnecessary.
•
Eliminate the RFI source.
10.9
Check for two-phase flow (slug flow)
Two-phase flow can cause rapid changes in the drive gain. This can cause a variety of
measurement issues.
1.
Check for two-phase flow alerts (e.g., A105).
If the transmitter is not generating two-phase flow alerts, two-phase flow is not the
source of your problem.
2.
Check the process for cavitation, flashing, or leaks.
3.
Monitor the density of your process fluid output under normal process conditions.
4.
Check the settings of
Two-Phase Flow Low Limit
,
Two-Phase Flow High Limit
, and
Two-Phase
Flow Timeout
.
Tip
You can reduce the occurrence of two-phase flow alerts by setting
Two-Phase Flow Low Limit
to a
lower value,
Two-Phase Flow High Limit
to a higher value, or
Two-Phase Flow Timeout
to a higher
value.
10.10
Check the drive gain
Excessive or erratic drive gain may indicate any of a variety of process conditions or sensor
problems.
To know whether your drive gain is excessive or erratic, you must collect drive gain data
during the problem condition and compare it to drive gain data from a period of normal
operation.
Excessive (saturated) drive gain
Possible causes and recommended actions for excessive (saturated)
drive gain
Table 10-5:
Possible cause
Recommended actions
Sensor tubes not completely
full
Correct process conditions so that the sensor tubes are full.
Plugged sensor tube
Check the pickoff voltages (see
). If either of them
are close to zero (but neither is zero), plugged tubes may be the
source of your problem. Purge the tubes. In extreme cases, you
may need to replace the sensor.
Drive board or module failure
Contact Micro Motion.
Troubleshooting
80
Micro Motion
®
LNG Series Meters
Summary of Contents for Micro Motion LNG Series
Page 1: ...Configuration and Use Manual MMI 20032809 Rev AA February 2017 Micro Motion LNG Series Meters ...
Page 6: ...Contents iv Micro Motion LNG Series Meters ...
Page 8: ...Getting Started 2 Micro Motion LNG Series Meters ...
Page 12: ...Before you begin 6 Micro Motion LNG Series Meters ...
Page 18: ...Quick start 12 Micro Motion LNG Series Meters ...
Page 20: ...Configuration and commissioning 14 Micro Motion LNG Series Meters ...
Page 50: ...Configure device options and preferences 44 Micro Motion LNG Series Meters ...
Page 56: ...Integrate the meter with the control system 50 Micro Motion LNG Series Meters ...
Page 58: ...Complete the configuration 52 Micro Motion LNG Series Meters ...
Page 60: ...Operations maintenance and troubleshooting 54 Micro Motion LNG Series Meters ...
Page 74: ...Measurement support 68 Micro Motion LNG Series Meters ...
Page 94: ...Using ProLink III with the transmitter 88 Micro Motion LNG Series Meters ...
Page 98: ...Default values and ranges 92 Micro Motion LNG Series Meters ...
Page 99: ...Default values and ranges Configuration and Use Manual 93 ...