Tip
Service port connections use standard connection parameters and a standard address. You do
not need to configure them here.
6.
Set the
PC Port
value to the PC COM port that you are using for this connection.
7.
Click
Connect
.
Need help?
If an error message appears:
•
Switch the leads and try again.
•
Ensure that you have specified the correct port on your PC.
•
Check the wiring between the PC and the transmitter.
A.1.3
Connect with ProLink III to the RS-485 port
You can connect directly to the RS-485 terminals on the transmitter or to any point on the
network.
Prerequisites
•
ProLink III installed and licensed on your PC
•
ProLink III v3.5 installed and licensed on your PC.
•
One of the following:
-
RS-232 to RS-485 signal converter
-
USB to RS-485 signal converter
•
An available serial port or USB port
•
Adapters as required (for example, 9-pin to 25-pin)
Procedure
1.
Attach the signal converter to the serial port or USB port on your PC.
2.
To connect over the RS-485 network:
a. Attach the leads from the signal converter to any point on the network.
b. Add resistance as necessary to achieve at least one volt across the connection
points.
3.
Start ProLink III.
4.
Choose
Connect to Physical Device
.
5.
Set the connection parameters to the values configured in the transmitter.
If your transmitter has not been configured, use the default values shown here.
Using ProLink III with the transmitter
86
Micro Motion
®
LNG Series Meters
Summary of Contents for Micro Motion LNG Series
Page 1: ...Configuration and Use Manual MMI 20032809 Rev AA February 2017 Micro Motion LNG Series Meters ...
Page 6: ...Contents iv Micro Motion LNG Series Meters ...
Page 8: ...Getting Started 2 Micro Motion LNG Series Meters ...
Page 12: ...Before you begin 6 Micro Motion LNG Series Meters ...
Page 18: ...Quick start 12 Micro Motion LNG Series Meters ...
Page 20: ...Configuration and commissioning 14 Micro Motion LNG Series Meters ...
Page 50: ...Configure device options and preferences 44 Micro Motion LNG Series Meters ...
Page 56: ...Integrate the meter with the control system 50 Micro Motion LNG Series Meters ...
Page 58: ...Complete the configuration 52 Micro Motion LNG Series Meters ...
Page 60: ...Operations maintenance and troubleshooting 54 Micro Motion LNG Series Meters ...
Page 74: ...Measurement support 68 Micro Motion LNG Series Meters ...
Page 94: ...Using ProLink III with the transmitter 88 Micro Motion LNG Series Meters ...
Page 98: ...Default values and ranges 92 Micro Motion LNG Series Meters ...
Page 99: ...Default values and ranges Configuration and Use Manual 93 ...