Notes
•
All fault alerts are displayed in the
Failed: Fix Now
category.
•
All information alerts are displayed in either the
Maintenance: Fix Soon
category or the
Advisory: Informational
category. The category assignment is hard-coded.
•
The transmitter automatically filters out alerts with
Alert Severity
set to
Ignore
.
2.
To acknowledge a single alert, check the
Ack
checkbox for that alert. To
acknowledge all alerts at once, click
Ack All
.
Postrequisites
•
To clear the following alerts, you must correct the problem, acknowledge the alert,
then power-cycle the transmitter: A001, A002, A010, A011, A012, A013, A029.
•
For all other alerts:
-
If the alert is inactive when it is acknowledged, it will be removed from the list.
-
If the alert is active when it is acknowledged, it will be removed from the list
when the alert condition clears.
Related information
Alert data in transmitter memory
8.5.2
Alert data in transmitter memory
The transmitter maintains three sets of data for every alert that is posted.
For each alert occurrence, the following three sets of data are maintained in transmitter
memory:
•
Alert List
•
Alert Statistics
•
Recent Alerts
Alert data in transmitter memory
Table 8-2:
Alert data structure
Transmitter action if condition occurs
Contents
Clearing
Alert List
As determined by the alert status bits, a list of:
•
All currently active alerts
•
All previously active alerts that have not
been acknowledged
Cleared and regenerated with every transmit-
ter power cycle
Alert Statistics
One record for each alert (by alert number)
that has occurred since the last master reset.
Each record contains:
•
A count of the number of occurrences
•
Timestamps for the most recent posting
and clearing
Not cleared; maintained across transmitter
power cycles
Transmitter operation
58
Micro Motion
®
LNG Series Meters
Summary of Contents for Micro Motion LNG Series
Page 1: ...Configuration and Use Manual MMI 20032809 Rev AA February 2017 Micro Motion LNG Series Meters ...
Page 6: ...Contents iv Micro Motion LNG Series Meters ...
Page 8: ...Getting Started 2 Micro Motion LNG Series Meters ...
Page 12: ...Before you begin 6 Micro Motion LNG Series Meters ...
Page 18: ...Quick start 12 Micro Motion LNG Series Meters ...
Page 20: ...Configuration and commissioning 14 Micro Motion LNG Series Meters ...
Page 50: ...Configure device options and preferences 44 Micro Motion LNG Series Meters ...
Page 56: ...Integrate the meter with the control system 50 Micro Motion LNG Series Meters ...
Page 58: ...Complete the configuration 52 Micro Motion LNG Series Meters ...
Page 60: ...Operations maintenance and troubleshooting 54 Micro Motion LNG Series Meters ...
Page 74: ...Measurement support 68 Micro Motion LNG Series Meters ...
Page 94: ...Using ProLink III with the transmitter 88 Micro Motion LNG Series Meters ...
Page 98: ...Default values and ranges 92 Micro Motion LNG Series Meters ...
Page 99: ...Default values and ranges Configuration and Use Manual 93 ...