Topics covered in this chapter:
•
Configure response time parameters
•
•
Configure informational parameters
5.1
Configure response time parameters
You can configure the rate at which process data is polled and process variables are
calculated.
Response time parameters include:
•
Update Rate
•
Calculation Speed
(
Response Time
)
5.1.1
Configure
Calculation Speed
(
Response Time
)
ProLink III
Device Tools
>
Configuration
>
Process Measurement
>
Response
>
Calculation Speed
Overview
Calculation Speed
is used to apply a different algorithm to the calculation of process variables
from the raw process data.
Calculation Speed
=
Special
produces faster and “noisier” response
to changes in the process.
Tip
You can use
Calculation Speed
=
Special
with either setting of
Update Rate
. The parameters control
different aspects of flowmeter processing.
Procedure
Set
Calculation Speed
as desired.
Option
Description
Normal
Transmitter calculates process variables at the standard speed.
Special
Transmitter calculates process variables at a faster speed.
Configure device options and preferences
Configuration and Use Manual
37
Summary of Contents for Micro Motion LNG Series
Page 1: ...Configuration and Use Manual MMI 20032809 Rev AA February 2017 Micro Motion LNG Series Meters ...
Page 6: ...Contents iv Micro Motion LNG Series Meters ...
Page 8: ...Getting Started 2 Micro Motion LNG Series Meters ...
Page 12: ...Before you begin 6 Micro Motion LNG Series Meters ...
Page 18: ...Quick start 12 Micro Motion LNG Series Meters ...
Page 20: ...Configuration and commissioning 14 Micro Motion LNG Series Meters ...
Page 50: ...Configure device options and preferences 44 Micro Motion LNG Series Meters ...
Page 56: ...Integrate the meter with the control system 50 Micro Motion LNG Series Meters ...
Page 58: ...Complete the configuration 52 Micro Motion LNG Series Meters ...
Page 60: ...Operations maintenance and troubleshooting 54 Micro Motion LNG Series Meters ...
Page 74: ...Measurement support 68 Micro Motion LNG Series Meters ...
Page 94: ...Using ProLink III with the transmitter 88 Micro Motion LNG Series Meters ...
Page 98: ...Default values and ranges 92 Micro Motion LNG Series Meters ...
Page 99: ...Default values and ranges Configuration and Use Manual 93 ...