Procedure
Set
Density Measurement Unit
to the option you want to use.
The default setting for
Density Measurement Unit
is
g/cm3
(grams per cubic centimeter).
Options for
Density Measurement Unit
The transmitter provides a standard set of measurement units for
Density Measurement Unit
.
Different communications tools may use different labels.
Options for
Density Measurement Unit
Table 4-6:
Unit description
ProLink III label
Specific gravity unit (not temperature-corrected)
SGU
Grams per cubic centimeter
g/cm3
Grams per liter
g/l
Grams per milliliter
g/ml
Kilograms per liter
kg/l
Kilograms per cubic meter
kg/m3
Pounds per U.S. gallon
lbs/Usgal
Pounds per cubic foot
lbs/ft3
Pounds per cubic inch
lbs/in3
API gravity
degAPI
Short ton per cubic yard
sT/yd3
4.5.2
Configure two-phase flow parameters
ProLink III
Device Tools
>
Configuration
>
Process Measurement
>
Density
Overview
The two-phase flow parameters control how the transmitter detects and reports two-
phase flow (gas in a liquid process or liquid in a gas process).
Note
Two-phase flow is sometimes referred to as
slug flow.
Procedure
1.
Set
Two-Phase Flow Low Limit
to the lowest density value that is considered normal in
your process.
Configure process measurement
Configuration and Use Manual
31
Summary of Contents for Micro Motion LNG Series
Page 1: ...Configuration and Use Manual MMI 20032809 Rev AA February 2017 Micro Motion LNG Series Meters ...
Page 6: ...Contents iv Micro Motion LNG Series Meters ...
Page 8: ...Getting Started 2 Micro Motion LNG Series Meters ...
Page 12: ...Before you begin 6 Micro Motion LNG Series Meters ...
Page 18: ...Quick start 12 Micro Motion LNG Series Meters ...
Page 20: ...Configuration and commissioning 14 Micro Motion LNG Series Meters ...
Page 50: ...Configure device options and preferences 44 Micro Motion LNG Series Meters ...
Page 56: ...Integrate the meter with the control system 50 Micro Motion LNG Series Meters ...
Page 58: ...Complete the configuration 52 Micro Motion LNG Series Meters ...
Page 60: ...Operations maintenance and troubleshooting 54 Micro Motion LNG Series Meters ...
Page 74: ...Measurement support 68 Micro Motion LNG Series Meters ...
Page 94: ...Using ProLink III with the transmitter 88 Micro Motion LNG Series Meters ...
Page 98: ...Default values and ranges 92 Micro Motion LNG Series Meters ...
Page 99: ...Default values and ranges Configuration and Use Manual 93 ...