Multi-measuring device for workplace analysis
BAPPU-
evo
8. Settings
In order to make adjustments to the following settings, select “Settings” in the main
menu.
8.1.
Language
The available languages can be called up using the
button. Subsequently, you can
select the language required.
8.2.
Display/Energy
Here you will find the display and energy functions where you can select the option
required with
. Using the
button, you can select the modes that are
available
8.3.
Measurement settings (measurement)
The following sections will explain how the settings to the individual measuring ranges
can be made.
Humidity compensation (rf-compensation)
Due to the fact that the relative humidity is dependent upon temperature, the
temperature of the housing will influence the values being measured; therefore, it is
important for the housing to have the same temperature as that of the ambient air.
For BAPPU, this means that when a cold measuring device (for example, when a
device has been left in the car) is brought into a heated room, the device will indicate
results that are much too high. This will be the case until the temperature of the device
housing is the same as that of the ambient air.
However, by implementing the temperature shock compensation on the humidity
sensor, the BAPPU-evo can be acclimatized rapidly.
In the default setting, the humidity compensation is active.
Time-weighting (time weighting noise)
Time weighting for the measurement values is distinguished by “slow” (1 second) and
“fast” (125 milliseconds). You can change between the time weighting during a
measurement by pressing “slow” or “fast” on the touch screen. The default setting is
“fast”.
page 25
status 07/2019