Multi-measuring device for workplace analysis
BAPPU-
evo
Tip:
Tip:
Before a new recording, the data in the counter
should be deleted in order to be able to get the maximum
recording time. You can see if there is any data in the
counter by clicking “Counter status”.
The “Logger status” button is used for calling up details pertaining to the status of the
memory content (existing long term recordings, data memory OK). “Erase Logger” is
used for deleting all of the data saved in the counter. The counter's internal clock is
also reset.
The data logger's “Selected measurement range” dialog box is used to select the
measurement range, whose measurement values are to be recorded during the
measurement.
The “Standard configuration” button automatically deselects the measurement range
noise level/min, display refresh rate and luminance density, because the recording of
these measurement areas is only of limited use as shown below.
Through specifying the memory data capture intervals, the user defines the
chronological gaps between two successive saved measurement values and thus the
averaging time period. This also affects the maximum amount of time available for the
recording. When the recording time increases, memory data capture intervals should
be increased in respect to memory resources and clarity.
Due to the fact that the minimum, maximum, and mean values of all selected
measurement ranges to be saved are recorded, it is possible to document the data of
individual measurement ranges with larger intervals.
“Store mode” is used to define whether a long-term recording is to be automatically
stopped once the counter memory is full or if the long-term recording should continue
(ring memory).
Data to be found at the beginning of the recording will be continually overwritten. For
example, this only makes sense when the end of the recording cannot be terminated
exactly, however, the current values are required.
Reading the BAPPU-evo data counter
To read the data, select “Logger/
Read Logger” in the menu list.
By clicking the “Connect Logger”
button, you can establish a
connection manually (an
automatic system connection
process is carried out by the
system every ten seconds).