Error message
Cause of the problem
Remedy
No printer or PC
found !
Connection failure between
gauge and printer or PC.
Switch on printer and/or check
the port setting of your PC.
Then, restart data transfer.
Data transfer
error !
Connection failure during data
transfer procedure.
Check the cable connection
and
the
settings
of
the
interface. Then, restart data
transfer.
Unsuitable substrate
(eg. magnetic)
Unsuitable substrate
(e.g. ferritic)
Substrate/coating
combination not suitable!
The calibration you have made
is not suitable for the shape or
substrate
of
measuring
sample.
The first two messages occur
in the Auto-FN mode, the other
one in the F or N mode.
Please carry out calibration on
an object which in shape and
substrate material is similar to
the
later
object
to
be
measured.
„Sensor-Problem !
Bitte wenden Sie
sich an den Service.“
Sensor failure ! Please contact
after-sales service.
Please
contact
after-sales
service.