Parameter “Calibration mode”
You can select from the following list:
Factory
Manual
ISO
SSPC
Rough
Australian
Swedish
With all predefined calibration methods (except for “Factory” and
“Manual”) the following message will appear once you have run through
the batch creation procedure: “Invalid calibration – recalibrate?”. At this
point you can decide whether to recalibrate immediately or at a later
point in measure mode.
Parameter “Substrate” (with FN sensors only)
When using the FN sensor, you can define a suitable substrate. The
following options are available:
Ferrous: magnetic induction principle for ferromagnetic substrates
(ferrous substrates, steel, alloy steel)
Non-ferrous: eddy currents principle for nonmagnetic, conducting
substrates (nonferrous substrates, austenitic steel)
Auto F/N: the sensor automatically identifies the substrate and selects
the matching measuring principle (this option is only possible with the
“Factory” or “Manual” calibration method.)
Parameter “Measuring mode”
Available options: “single reading” and “Scan mode”.
In the “single readings” mode, a reading is taken and stored into the
batch each time you put the sensor onto the sample, i.e. put the sensor
on the sample to take a reading, lift the sensor and put the sensor down
again to take the next one.
In the “Scan mode”, readings are taken and displayed continuously as
soon as you put the sensor on the sample. Lift the sensor to stop taking
readings. Press OK to store the latest single reading to the batch.
Parameter “Roughness”
To compensate for roughness, you can enter a correction value
“roughness” as related to the current roughness value of your substrate.