15. A
© ElektroPhysik
MiniTest 725, MiniTest 735, MiniTest 745
78 / 85
Error message
Cause of the problem
Remedy
Check sensor
connection !
Gauge sensor interrupt during
operation.
Possible causes:
The external sensor has been
disconnected during operation.
or
The internal sensor has been
disconnected during operation
through
-
loose sensor
connection
-
defective cable
-
defective sensor
Please for check reason of
error.
If the sensor seems to be
defective, please exchange by
another one or contact the
ElektroPhysik
after-sales
service.
No sensor
found !!!
After switch-on, the gauge
cannot establish a connection
to the sensor.
Possible causes:
-
no sensor connected
-
loose sensor
connection
-
defective cable
-
defective sensor
Please check for reason of
error.
When using MiniTest 745:
Please try to use the senor in
the internal mode. If it works in
this mode, the sensor cable is
the cause of error.
Please replace sensor cable.
If a defective sensor is the
cause of error, please replace
by
new
one
or
contact
ElektroPhysik
after
sales
service.