6.4.3.4 Multi-point calibration
Start calibration and put the sensor on the blank or uncoated calibration
sample. Wait for the signal to sound and lift sensor.
Repeat this procedure several times (approx. 3 to 10 times) on the
same measuring spot. The average
will be shown. Evaluation of
average is to increase accuracy of calibration. Accuracy will increase
with an increasing number of readings. Press OK to confirm the
calibration point. Press OK once again to complete the calibration
procedure.
Put the first precision standard on the uncoated calibration sample.
Place the sensor on top of it, wait for the signal to sound and lift sensor.
Repeat this procedure several times (approx. 3 to 10 times) on the
same measuring spot. The average will be shown.
If the shown set point (Cal) as displayed is not identical with the
thickness of your precision standard, use arrow up and down keys to
adjust accordingly.
Press OK to confirm the calibration point.
Put the second precision standard on the uncoated calibration sample.
Place the sensor on top of it, wait for the signal to sound and lift sensor.
Repeat this procedure several times (approx. 3 to 10 times) on the
same measuring spot. The average will be shown.
If the shown set point (Cal) as displayed is not identical with the
thickness of your precision standard, use arrow up and down keys to
adjust accordingly.
Press OK to confirm the calibration point.