Action
Step
Tap OFFSETS on the local control software home screen.
3
Result: The Transmitter Offsets screen opens.
Type the calculated offsets into the Offset text boxes.
4
Tap HOME to return to the home screen.
5
If the adjustment of the temperature probe at the operating temperature does not bring
the system into tolerance, replace the temperature probe and repeat the calibration
procedure.
If the calibration of the new temperature probe does not bring the system into tolerance,
contact your GE representative for service.
290
Xcellerex XDUO 2500 Mixer Operating Instructions 29146667 AH
10 Reference information
10.5 Temperature probe calibration